Determination of 7 pyrethroids residues in aquatic product by gas chromatography
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Gas chromatography with electron capture detection method was developed for the determination of 7 pyrethroids residues in aquatic products. Pyrethroids were extracted with cyclohexane and ethylacetate(1:1,V/V)and lipids were removed with petroleum ether. Cleanup was performed on a LC-Florisil column. The working curves for the pyrethroids were linear (r>0.992) in the range of 1 to 100 μg/L. When spiked with 5, 10 and 20 μg/kg of pyrethroids, the average recoveries were 70.0%~114.8% and the intra-day RSD(n=5) were 1.2%~9.8% and the inter-day RSD(n=3) were 3.8%~11.2%. The detection limit as estimated by 2 times signal to noise method for bifenthrin, fenpropathrin, lambdacyhalothrin, permethrin was 1 μg/kg and for permethrin, fenvalerate, deltamethrin was 2 μg/kg respectively.

    Reference
    Related
    Cited by
Get Citation

GAO Yu-yi, SHEN Mei-fang, SONG Hong-bo, WU Guanghong, GE Xiao-qin, GENG Xue-bing. Determination of 7 pyrethroids residues in aquatic product by gas chromatography[J]. Journal of Fisheries of China,2009,33(1):132~138

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:February 28,2008
  • Revised:May 08,2008
  • Adopted:September 18,2008
  • Online: January 08,2009
  • Published: